Investigation of the Temperature Effect of Ultrasound Used in Cancer Therapy
Yazarlar/Oluşturanlar
- 1. TUBITAK UME Med Metrol Lab, Baris Mah Dr Zeki Acar Cad 1, TR-41470 Gebze, Kocaeli, Turkey
Açıklama
In present study, temperature rise in a tissue mimicking material induced by a HIFU (High Intensity Focused Ultrasound) transducer was examined with two different methods. In order to determine the temperature rise inside and distribution on the surface of a Tissue-Mimicking Material (TMM), T-type thermocouple (TC) temperature sensors and infrared (IR) camera system was used in the first and second method respectively. During the ultrasonic power application, temperature increase and distribution in the TMM have been represented. Temperature measurements were realized inside a TMM which has characteristics very similar to human liver tissue. Temperature rise has been measured for different input ultrasonic powers and durations. Two different temperature measuring tools were compared and consistency between these methods was presented.